Page 84 © Boule Medical AB, March 2018. Article no. 1504496
9. Troubleshooting and System Messages
Sample Pathology Messages and Flag Indicators
HCT
Indicator/Message Criteria Description Action
Pathological
Message
If HCT is > 10%
below lower limit
HCT: Anemia; evaluate RBC on slide Blood sample too old
or pathological sample.
Re-analyze sample, if
still agged an alternative
methodology such as e.g.
slide review, is adviced.
Pathological
Message
If HCT is > 10%
above upper limit
HCT: Evaluate patient for causes of
polycythemia
MCHC
Indicator/Message Criteria Description Action
Pathological
Message
If MCHC is >
10% below lower
limit
MCHC: In the following order:
Evaluate for extreme RBC regeneration
Run Control
Blood sample too old
or pathological sample.
Re-analyze sample, if
still agged an alternative
methodology such as e.g.
slide review, is adviced.
Pathological
Message
If MCHC is >
10% above upper
limit
MCHC: In the following order:
Evaluate for turbidity, lipemia, and
extreme hemolysis
Heinz bodies – cat
Evaluate for agglutination /spin crit
Run Control
PLT
Indicator/Message Criteria Description Action
Pathological
Message
If PLT is > 25%
below lower limit
PLT: Evaluate platelets on slide Blood sample too old
or pathological sample.
Re-analyze sample, if
still agged an alternative
methodology such as e.g.
slide review, is adviced.
Pathological
Message
If PLT is > 50%
above upper limit
PLT: Evaluate histogram for extreme RBC
microcytosis
Prime Cycle
The prime cycle is used to reset the analyzer after an error has been indicated or a failure in running