NISEA
2.2 Main features
The vacuum pump ensures an optimal sterilisation process in which the steam interacts properly with the load being
processed,includinganycavities.
2.2.1 Sterilisation programs
● Type B
-134°CStandard
-134°CPrion
-134°CB-Flash
-121°CStandard
● Special cycle
- Drying only
● Type S
-134°CUniversal S
-134°CSoftprogramm
2.2.2 Test programs
- Vacuum test: checks the ability to hold a vacuum
- Bowie & Dick test: checks steam penetration into porous loads
- 121 or 134°C Helix test: checks steam penetration into hollow loads
Theautoclaveisavailablewiththefollowingstoragedevices:
- USB port (USB peripheral not supplied).
-RS-232port(forFaroprinter,NOTsupplied).
2.2.3 Water lling systems
Automatic: with the aid of a pump inside the autoclave.
Mains connectionpossibilityofconnectingademineralisationsystem,toincreaseautonomy.Itmustbeinstalledby
an authorised technician.
Manual:asanalternative,thedeviceisequippedwithamanualemergencyllingsystemattheFRONT(9 g. 1).
2.2.4 Automatic sterilisation parameter control
Duringsterilisation,thepressure,temperatureandtimearecontinuouslymonitoredbyanautomaticprocesscontrol
system.
2.2.5 Motorised closure
The autoclave has a motorised closure system to ensure a good seal. It is based on a screw and nut screw. Closure
works with a pre-locking mode that activates when the door is shut. The door is only closed fully when the cycle starts.
2.2.6 Bio-X lter
Thisinjectssterileairintothechamberduringthedryingandpressurebalancingphases.
2.1 Variants
The following autoclave variants are available:
Version Basic
Chamber size
18 23
Power supply
REF Cycles B
750 030XXX 750 000XXX 750 020XXX 750 130XXX 750 100XXX 750 120XXX
REF Cycles B Switzerland
// 750 010XXX // // 750 110XXX //
REF Cycles S
750 031XXX 750 001XXX 750 021XXX 750 131XXX 750 101XXX 750 121XXX
REF Cycles S
Switzerland
// 750 011XXX // // 750 111XXX //
REF Cycles B + S
750 032XXX 750 002XXX 750 022XXX 750 132XXX 750 102XXX 750 122XXX
REF Cycles B + S Switzerland
// 750 012XXX // // 750 112XXX //
Version Premium
Chamber size
18 23
Power supply
REF Cycles B
751 030XXX 751 000XXX 751 020XXX 751 130XXX 751 100XXX 751 120XXX
REF Cycles B Switzerland
// 751 010XXX // // 751 110XXX //
REF Cycles S
751 031XXX 751 001XXX 751 021XXX 751 131XXX 751 101XXX 751 121XXX
REF Cycles S
Switzerl
// 751 011XXX // // 751 111XXX //
REF Cycles B + S
751 032XXX 751 002XXX 751 022XXX 751 132XXX 751 102XXX 751 122XXX
REF Cycles B + S Switzerland
// 751 012XXX // // 751 112XXX //
XXX:
they mean an aesthetic personalization or a thermodynamic cycle already validated.
220 V 230 V 240 V 220 V 230 V 240 V
60 Hz 50/60 Hz 50 Hz 60 Hz 50/60 Hz 50 Hz
220 V 230 V 240 V 220 V 230 V 240 V