6 TROUBLESHOOTING GUIDE
6.1 Error Codes
6.2 Alarm Codes
6 TROUBLESHOOTING GUIDE
6.1 Error Codes
Code Message Description
Reset
method
ER 001 Door closing timeout The door closing time is too long B
ER 002 Door switch Door switch error condition B
ER 003 Door unlocking timeout The door unlocking time is too long B
ER 004 Door opening timeout The door opening time is too long B
ER 005 Poor water quality The water conductivity is higher than 10 uS/cm B
ER 006 Very poor water quality The water conductivity is higher than 20 uS/cm B
ER 100 Preheating timeout The vacuum preheating phase time is too long A
ER 101 Preheating timeout The vacuum preheating phase time is too long A
ER 110 Vacuum pressure timeout The time to reach the vacuum pressure is too long A
ER 120 Discharge pressure timeout The time to reach the thermodynamic discharge pressure is too long A
ER 121 Pressure difference The pressure increase is lower than the set parameter A
ER 130 Discharge time The pressure discharge time is too long A
ER 140 Sterilisation pressure timeout The time to reach the sterilisation pressure is too long A
Code Message Description
Reset
method
ER 141 Pressure difference The pressure increase is lower than the set parameter A
ER 150 T chamber high The chamber temperature is over the top limit for sterilisation A
ER 151 T chamber low The chamber temperature is below the bottom limit for sterilisation A
ER 152 T theoretical high The theoretical temperature is over the top limit for sterilisation A
ER 153 T theoretical low The theoretical temperature is below the bottom limit for sterilisation A
ER 154 Temperature difference
The difference between the chamber and theoretical temperatures is over the
allowed limit
A
ER 160 Final discharge The final discharge time is too long A
ER 170 AV block The measured pressure is below the set parameter for the drying phase A
ER 180 Balancing timeout The pressure balancing time is too long A
ER 200 Vacuum test timeout The time to reach the vacuum pressure is too long in the vacuum test A
ER 201 Vacuum Test: first phase Failure to comply with the parameters in the first phase of the vacuum test A
ER 202 Vacuum Test: second phase Failure to comply with the parameters in the second phase of the vacuum test A
ER 203 Vacuum Test: temperature Failure to comply with the temperature limits during the vacuum test A
ER 204 Vacuum Test: balancing Pressure balancing failed in the vacuum test A
6.2 Alarm Codes
Code Message Description
Reset
method
AL 001 Reset Pressed The reset button was pressed A
AL 002 Blackout A blackout has occurred A
AL 003 Water filling time The time to fill the feed tank is too long B
AL 100 Chamber probe fault The chamber probe reading circuit is faulty A/B
AL 101 Generator Probe Fault The generator probe reading circuit is faulty A/B
AL 102 Bundle probe fault The pipe bundle probe reading circuit is faulty A/B
AL 103 Pressure transducer The pressure transducer is faulty A/B
AL 104 Door microswitch The door microswitch has opened unexpectedly A
AL 105 Safety microswitch The safety microswitch has opened unexpectedly A
AL 200 Chamber safety cut-off The chamber has exceeded the safety temperature A
AL 201 Generator safety cut-off The generator has exceeded the safety temperature A
AL 202 Bundle safety cut-off The pipe bundle has exceeded the safety temperature A
AL 203 Safety valve fault The pressure has exceeded the maximum trip threshold of the safety valve A
AL 300 Mains not synchronised
There is no synchronisation for managing AC loads and checking the mains
frequency.
A
AL 301 Valve drive voltage There is no valve drive voltage due to a PTC trip A
Reset
procedure
Reset
procedure
6 GUIDA AI PROBLEMI
6.1 Codici Errori
Codice Messaggio Descrizione
Modali
reset
ER 001 Timeout chiusura portello Superamento del tempo per la chiusura del portello B
ER 002 Switch portello Condizione di errore dello switch portello B
ER 003 Timeout sblocco portello Superamento del tempo per lo sblocco del portello B
ER 004 Timeout apertura portello Superamento del tempo per l'apertura del portello B
ER 005 Qualità acqua scarsa La conducibilidell'acquaèsuperioreai15uS/cm B
ER 006 Qualità acqua pessima La conducibilità dell'acqua è superiore ai 0uS/cm B
ER 100 Timeout preriscaldo Superamento del tempo per le fasi di vuoto in preriscaldo A
ER 101 Timeout preriscaldo Superamento del tempo per le fasi di salita pressione in preriscaldo A
ER 110 Timeout pressione del vuoto Superamento del tempo per il raggiungimento della pressione di vuoto A
ER 120 Timeout pressione di scarico
Superamento del tempo per il raggiungimento della pressione dello scarico
termodinamico
A
ER 121 Delta pressione Incremento di pressione inferiore al parametro stabilito A
ER 130 Timeout scarico Superamento del tempo per lo scarico della pressione A
3
6 GUIDA AI PROBLEMI
6.1 Codici Errori
Codice Messaggio Descrizione
Modali
reset
ER 001 Timeout chiusura portello Superamento del tempo per la chiusura del portello B
ER 002 Switch portello Condizione di errore dello switch portello B
ER 003 Timeout sblocco portello Superamento del tempo per lo sblocco del portello B
ER 004 Timeout apertura portello Superamento del tempo per l'apertura del portello B
ER 005 Qualità acqua scarsa La conducibilidell'acquaèsuperioreai15uS/cm B
ER 006 Qualità acqua pessima La conducibilità dell'acqua è superiore ai 0uS/cm B
ER 100 Timeout preriscaldo Superamento del tempo per le fasi di vuoto in preriscaldo A
ER 101 Timeout preriscaldo Superamento del tempo per le fasi di salita pressione in preriscaldo A
ER 110 Timeout pressione del vuoto Superamento del tempo per il raggiungimento della pressione di vuoto A
ER 120 Timeout pressione di scarico
Superamento del tempo per il raggiungimento della pressione dello scarico
termodinamico
A
ER 121 Delta pressione Incremento di pressione inferiore al parametro stabilito A
ER 130 Timeout scarico Superamento del tempo per lo scarico della pressione A