6.4 Reset procedures
Reset procedure ”A”
Ifthecycleis“running”,resettheerrorbyselectingtheconrmicononthesafetyscreenandthenselectthe
Home icon to return to the initial screen.
Reset procedure ”B”
Resettheerrorbeselectingtheconrmicononthesafetyscreen.
N.B. In the event of faults that compromise sterilisation/test cycle execution, the cycle selection icons on
the Home screen will be disabled.
6.5 Troubleshooting
Error reset procedure “B”
Reset the error be selecting the confirm button on the safety screen.
N.B. In the event of faults that compromise sterilisation/test cycle execution, the cycle selection icons on the Home screen will be
disabled.
6.5 Troubleshooting
Code Possible cause Possible solution
ER 001
There is a malfunction in the motorised door
closing/opening system.
Reset the machine.
Repeat the door closing procedure.
If the problem persists, contact technical support.
ER 002
There is a malfunction in the motorised door
closing/opening system.
Reset the machine.
Repeat the door closing procedure.
If the problem persists, contact technical support.
ER 003
There is a malfunction in the motorised door
closing/opening system.
Reset the machine.
Repeat the door opening procedure.
If the problem persists, contact technical support.
ER 004
There is a malfunction in the motorised door
closing/opening system.
Reset the machine.
Repeat the door opening procedure.
If the problem persists, contact technical support.
ER 005 The water quality is poor.
Reset the machine.
Poor water quality: change the water in the feed tank with better
quality water.
ER 006 The water quality is very poor.
Reset the machine.
Very poor water quality: empty the feed tank and refill it with better
quality water.
If the problem persists, contact technical support.
6.3 Messagges
Codice
Fase Messaggio
Msg 001 Standby o avvio ciclo Serbatoio di carico vuoto
Msg 002 Standby o avvio ciclo Serbatoio di scarico pieno
Msg 003 Inserimento Pin Pin errato
Msg 004 Inserimento Pin Inserire Pin
Msg 005 Avvio Ciclo Porta aperta
Msg 006 Trasferimento dati su USB Trasferimento dati non riuscito
Msg 007
Termine positivo ciclo test valvola
sicurezza
Pressione di attivazione valvola sicurezza P=xxxkPa
Msg 008 Avvio Vuoto Test Temperatura di camera oltre il limite
Msg 009 Accensione macchina Serial number incongruenti fra CPU e MB
Msg 010 Accensione macchina Eseguire lubrificazione sistema motorizzato porta
Msg 011 Accensione macchina Sostituire filtro Bio-X
Msg 012 Accensione macchina Eseguire manutenzione
Pag.64
6.6 Risoluzione messaggi
Codice Fase Messaggio Possibile soluzione
Msg 001 Standby o avvio ciclo Serbatoio di carico vuoto Riempire il serbatoio di carico
Msg 002 Standby o avvio ciclo Serbatoio di scarico pieno Svuotare il serbatoio di scarico
Msg 003 Inserimento Pin Pin errato Introdurre il Pin corretto
Msg 004 Inserimento Pin Inserire Pin Inserire Pin
Msg 005 Avvio Ciclo Porta aperta Chiudere la porta
Msg 006 Trasferimento dati su USB Trasferimento dati non riuscito Verificare il supporto di memoria e ripetere l’operazione
Msg 007
Termine positivo ciclo test
valvola sicurezza
Pressione di attivazione valvola
sicurezza P=xxxkPa
-----
Msg 008 Avvio Vuoto Test
Temperatura di camera oltre il
limite
Ripetere il Vuoto test quando la temperatura di camera
rientra nel limite ammesso (45°C).
Msg 009 Accensione macchina
Serial number incongruenti fra
CPU e MB
Contattare il Servizio tecnico Faro
Msg 010 Accensione macchina
Eseguire lubrificazione sistema
motorizzato porta
Eseguire la lubrificazione del sistema motorizzato secondo le
istruzioni al par. 5.1.9 del manuale d’uso.
Msg001
Msg002
Msg003
Msg004
Msg005
Msg006
Msg007
Msg008
Msg009
Msg010
Msg011
Msg012
Feed tank empty
Feed tank full
Incorrect PIN
Enter PIN
Door open
Transfer not succssful
Safety valve activation pressure P=xxx kPa
Chamber temperature over limit
Serial Number inconsistency between CPU and MB
Lubricate the automated system of the door
Replace the BIO-X lter
Conduct maintenance
Standby or cycle start
Standby or cycle start
PIN entry
PIN entry
Cycle start
Data transfer to USB
Safety valve test cycle successful
Vacuum Test Start
Machine switch-on
Machine switch-on
Machine switch-on
Machine switch-on
Code
Phase
Message
AL 402 Band error Band failure
Code Message Description
Reset
method
AL 302 Mains frequency error The mains frequency is < 45Hz or > 65Hz A
AL 303 Watchdog trip Program execution has crashed A
AL 304 Communication error with CPU There is a communication fault between the main board and the CPU board A
AL 305 Hardware error Hardware fault on the main board A
AL 306 Hardware error
Hardware fault on the main board
A
AL 307 Hardware error
Hardware fault on the main board
A
AL 308 Hardware error
Hardware fault on the main board
A
AL 309 Hardware error
Hardware fault on the main board
A
AL 310 Hardware error
Hardware fault on the main board
A
Reset
procedure
Code Message Description
Reset
method
AL 302 Mains frequency error The mains frequency is < 45Hz or > 65Hz A
AL 303 Watchdog trip Program execution has crashed A
AL 304 Communication error with CPU There is a communication fault between the main board and the CPU board A
AL 305 Hardware error Hardware fault on the main board A
AL 306 Hardware error
Hardware fault on the main board
A
AL 307 Hardware error
Hardware fault on the main board
A
AL 308 Hardware error
Hardware fault on the main board
A
AL 309 Hardware error
Hardware fault on the main board
A
AL 310 Hardware error
Hardware fault on the main board
A
6.2 Alarm Codes
Code Message Description
Reset
method
AL 001 Reset Pressed The reset button was pressed A
AL 002 Blackout A blackout has occurred A
AL 003 Water filling time The time to fill the feed tank is too long B
AL 100 Chamber probe fault The chamber probe reading circuit is faulty A/B
AL 101 Generator Probe Fault The generator probe reading circuit is faulty A/B
AL 102 Bundle probe fault The pipe bundle probe reading circuit is faulty A/B
AL 103 Pressure transducer The pressure transducer is faulty A/B
AL 104 Door microswitch The door microswitch has opened unexpectedly A
AL 105 Safety microswitch The safety microswitch has opened unexpectedly A
AL 200 Chamber safety cut-off The chamber has exceeded the safety temperature A
AL 201 Generator safety cut-off The generator has exceeded the safety temperature A
AL 202 Bundle safety cut-off The pipe bundle has exceeded the safety temperature A
AL 203 Safety valve fault The pressure has exceeded the maximum trip threshold of the safety valve A
AL 300 Mains not synchronised
There is no synchronisation for managing AC loads and checking the mains
frequency.
A
AL 301 Valve drive voltage There is no valve drive voltage due to a PTC trip A
Codice Messaggio Descrizione
Modalità
reset
AL 300 Mancanza sincronismo rete
Assenza di sincronismo per la gestione carichi AC e verifica della frequenza
di rete.
A
AL 301 Tensione comando valvole Mancanza tensione di comando valvole per intervento PTC A
AL 302 Errore frequenza rete elettrica Frequenza rete elettrica < 45Hz o > 65Hz A
AL 303 Intervento watchdog Esecuzione codice in blocco A
AL 304 Errore di comunicazione con CPU Guasto di comunicazione tra scheda principale e scheda CPU A
AL 305 Errore hardware Guasto hardware sulla scheda principale A
AL 306 Errore hardware
Guasto hardware sulla scheda principale
A
AL 307 Errore hardware
Guasto hardware sulla scheda principale
A
AL 308 Errore hardware
Guasto hardware sulla scheda principale
A
AL 309 Errore hardware
Guasto hardware sulla scheda principale
A
AL 310 Errore hardware
Guasto hardware sulla scheda principale
A
AL 401 Generator error
Generator failure
AL 402 Band error Band failure