5 ROUTINE MAINTENANCE
5.1 General table
Beforecleaninganysurface,makesurethatithascooleddown.
115115
NISEA
5 ROUTINE MAINTENANCE
5.1 General table
Frequency
Operation
Daily
5 cycles
Monthly
50 cycles
Quarterly
500 cycles
Annually
1000
cycles
Clean the porthole seal ●
Clean the chamber edge (area
where the seal touches)
●
Clean the porthole ●
Clean the chamber and accessories ●
Clean the front panel ●
Clean the side walls and top panel ●
Clean the chamber filter ●
Clean the dust filter (if provided) ●
Clean the screw and nut screw ●
Lubricate the screw and nut screw ●
●
Replace the bacterial filter ●
Replace the porthole seal ●
Before cleaning any surface, make sure that it has cooled down
5.1.1 Cleaning the sterilisation chamber and internal components
Use a soft cloth soaked in water to avoid damaging the protective surface treatment of the
chamber. Rinse thoroughly with water and then dry. If necessary, moisten the cloth with white
vinegar to remove any limescale.
However, we recommend cleaning the chamber only when strictly necessary. If there are
limescale deposits, immediately check what is causing them. White patches forming at the
bottom of the chamber means that the water used is of poor quality or does not comply with the
required specifications.
5.1.2 Cleaning the porthole
Use a soft cloth soaked in water to avoid damaging the protective treatment of the porthole.
Rinse thoroughly with water and then dry. DO NOT USE metallic or particularly abrasive
cloths for cleaning. Using them could damage the external surface of the porthole.