Code Possible cause Possible solution
ER 100
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
Reset the machine.
Check that the door seal is clean and positioned correctly, then repeat
the selected program.
If the problem persists, contact technical support.
ER 101 There is a leak in the machine plumbing circuit.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 110
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 120
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
ER 121
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 130 There is a malfunction in the machine discharge system.
Reset the machine.
Check the amount of load in the chamber and repeat the selected
program.
If the problem persists, contact technical support.
ER 140
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
Reset the machine.
Check:
Code Possible cause Possible solution
ER 141
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 150
There is a malfunction in the machine discharge system
or a chamber loading error.
Reset the machine.
Check:
a) the tray support position and the tray positions in the chamber;
b) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 151
There is a malfunction in the sterilisation condition
holding system or a chamber loading error.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 152
There is a malfunction in the sterilisation condition
holding/control system
ER 153
There is a malfunction in the sterilisation condition
holding/control system
Reset the machine.
Check:
a) that there is water in the feed tank;
b) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 154
There is a malfunction in the sterilisation condition
holding/control system
Reset the machine.
Check:
a) the tray support position and the tray positions in the chamber;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
Code Possible cause Possible solution
If the problem persists, contact technical support.
ER 160 There is a malfunction in the machine discharge system
Reset the machine.
Check the amount of load in the chamber and then repeat the selected
program.
If the problem persists, contact technical support.
ER 170 There is a malfunction in the machine drying system
ER 180 There is a malfunction in the pressure balancing system
Reset the machine.
Check the amount of load in the chamber and the Bio-X filter
condition, then repeat the selected program.
If the problem persists, contact technical support.
ER 200
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
ER 201 There is a leak in the machine plumbing circuit.
ER 202 There is a leak in the machine plumbing circuit.
Reset the machine.
Check that the door seal is clean and positioned correctly, then repeat
the selected program.
If the problem persists, contact technical support.
ER 203
A Vacuum Test is running in chamber conditions other
than ambient temperature.
Reset the machine.
Repeat the Vacuum Test, making sure that the chamber is at ambient
temperature.
If the problem persists, contact technical support.
ER 204 There is a malfunction in the pressure balancing system
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
AL 001 The reset button was pressed Reset the machine.