FEASA LED ANALYSER
ICT VERSION
Capture Sequence Mode
IdentifySequence - Identify the current sequence pattern
Transmit Receive
identifysequence ##
Where:
## represents the number of a stored sequence pattern and is a value in the range 00
– 20 which compares to the current sequence capture executed.
Description
This command compares the current sequence pattern with a stored list of patterns and
identifies it. The pattern is identified by a number in the range 00 – 20. Up to 21 patterns can
be stored. This feature allows the Analyser to identify patterns by comparing them against
patterns stored in memory.
The setsequencemode#, setsequencechannel@@, setsequencenumber###,
sequencecapture###, sequencenumber, sequencethreshold#####,
sequencestartdelay### and sequencewait### commands must be executed before this
command takes effect. Also, a capturesequence command must be issued before the
identifysequence command.
Example:
The PC transmits identifysequence to the LED Analyser to instruct it to identify the current
pattern and return its identity number.
identifysequence
04
Back to Index
120

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Feasa ICT Series and is the answer not in the manual?

Feasa ICT Series Specifications

General IconGeneral
BrandFeasa
ModelICT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals