FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 21 of 170
4.6.3 RS232C Serial Port
FT-112(D) Panel weighing indicator has RS232C serial port which is galvanically isolated from other circuitry to
increase the EMC immunity. The usage of this serial port and its specifications are described in the Table 4.3
and its pin configuration is shown in Table 4.4.
Interfacing with printer, PC, PLC, remote display etc.
Continuous, Fast Continuous, Printer, BSI Protocol or Modbus RTU
1200 / 2400 / 4800 / 9600 (Default) / 19200 / 38400 / 57600 bps
Even, Odd or No (default)
1 start bit and 1 stop bit
Table 4.3 - RS 232C Serial Interface Specifications
Table 4.4 – Pin configuration of RS232C terminals
2 wire connection to peripherals: RS 232C serial connection is done with two wires as indicated below in
Figure 4.7. if there is no data entry to the weighing indicator. Printer, remote display etc. connections can be
done with 2 wire.
Figure 4.7 – 2 wire RS 232C connection with a printer or PC
3 wire connection to the peripherals: RS232C serial connection is done with three wires as indicated below
in Figure 4.8 for bidirectional interfacing. Typical application is bidirectional BSI format interfacing with PC or
PLC.
Figure 4.8 – 3 wire RS232C connection with a PC or PLC