FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 41 of 170
6 PROGRAMMING AND CALIBRATION
You will find the programming and calibration procedure of FT-112(D) Panel weighing indicator in this section.
The arrow on navigation keys indicates the function of the keys in programming menu. The basic meanings of
these keys are;
A set-up DIP switch is located near the load cell connector at the FT-112 (D) panel rear side as shown in the
figure below and its position should be “ON” (downward) to change the metrological related parameters
including calibration.
Figure 6.1- The location of calibration DIP switch.
Entering to the Programming and Calibration
Enter the programming menu as described below. Legally related parameters can be changed if the
calibration dip switch is at ON position. These parameters are marked with M symbol in the parameter table in
this section. If you will change any legally relate parameter or will perform calibration, set the calibration switch
to the ON position before entering to the programming.