34
This section presents a compilation of emissivity data from the infrared literature and
measurements made by FLIR Systems.
34.1 References
1. Mikaél A. Bramson: Infrared Radiation, A Handbook for Applications, Plenum press,
N.Y.
2. William L. Wolfe, George J. Zissis: The Infrared Handbook, Office of Naval Research,
Department of Navy, Washington, D.C.
3. Madding, R. P.: Thermographic Instruments and systems. Madison, Wisconsin: Uni-
versity of Wisconsin Extension, Department of Engineering and Applied Science.
4. William L. Wolfe: Handbook of Military Infrared Technology, Office of Naval Research,
Department of Navy, Washington, D.C.
5. Jones, Smith, Probert: External thermography of buildings..., Proc. of the Society of
Photo-Optical Instrumentation Engineers, vol.110, Industrial and Civil Applications of
Infrared Technology, June 1977 London.
6. Paljak, Pettersson: Thermography of Buildings, Swedish Building Research Institute,
Stockholm 1972.
7. Vlcek, J: Determination of emissivity with imaging radiometers and some emissivities
at λ = 5 µm. Photogrammetric Engineering and Remote Sensing.
8. Kern: Evaluation of infrared emission of clouds and ground as measured by weather
satellites, Defence Documentation Center, AD 617 417.
9. Öhman, Claes: Emittansmätningar med AGEMA E-Box. Teknisk rapport, AGEMA
1999. (Emittance measurements using AGEMA E-Box. Technical report, AGEMA
1999.)
10. Matteï, S., Tang-Kwor, E: Emissivity measurements for Nextel Velvet coating 811-21
between –36°C AND 82°C.
11. Lohrengel & Todtenhaupt (1996)
12. ITC Technical publication 32.
13. ITC Technical publication 29.
14. Schuster, Norbert and Kolobrodov, Valentin G. Infrarotthermographie. Berlin: Wiley-
VCH, 2000.
Note The emissivity values in the table below are recorded using a shortwave (SW)
camera. The values should be regarded as recommendations only and used with
caution.
34.2 Tables
Table 34.1 T: Total spectrum; SW: 2–5 µm; LW: 8–14 µm, LLW: 6.5–20 µm; 1: Material; 2: Specification;
3:Temperature in °C; 4: Spectrum; 5: Emissivity: 6:Reference
1 2 3 4 5 6
3M type 35 Vinyl electrical
tape (several
colors)
< 80 LW 0.96 13
3M type 88 Black vinyl electri-
cal tape
< 105 LW 0.96 13
3M type 88 Black vinyl electri-
cal tape
< 105 MW < 0.96 13
3M type Super 33
+
Black vinyl electri-
cal tape
< 80 LW 0.96 13
Aluminum anodized sheet 100 T 0.55 2
Aluminum anodized, black,
dull
70
SW
0.67 9
Aluminum anodized, black,
dull
70 LW 0.95 9
Aluminum anodized, light
gray, dull
70
SW
0.61 9
#T559845; r. AJ/37554/37554; en-US
162

Table of Contents

Other manuals for FLIR E60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR E60 and is the answer not in the manual?

FLIR E60 Specifications

General IconGeneral
Resolution320 x 240 pixels
Thermal Sensitivity<0.05°C at 30°C
Field of View25° x 19°
Battery Life4 hours
Spatial Resolution1.36 mrad
Image Frequency60 Hz
MeterLinkYes
Laser PointerYes
Text AnnotationYes
Storage MediaSD card
Spectral Range7.5 – 13 µm
Temperature Range-20°C to +650°C
Display3.5" LCD touchscreen
FocusManual
Accuracy±2°C or ±2% of reading
Image ModesThermal, visual, MSX
Voice AnnotationYes
Battery TypeLi-Ion battery
Minimum Focus Distance0.4 m
Operating Temperature Range-15°C to +50°C
Storage Temperature Range-40°C to +70°C
ConnectivityWi-Fi, USB