Screening alarm
21
21.1 General
The screening alarm can be used, for example, at airports to detect passengers with ele-
vated body temperatures, which may indicate the presence of a fever.
Activating the screening alarm will turn on a measurement box and screening data in the
result table.
The sampled average temperature.
The alarm temperature.
The measured temperature.
The alarm will trigger when the measurement box measures a temperature higher than
the alarm temperature. The alarm temperature is, in turn, the sum of a specified allowed
deviation and a sampled average value.
21.2 Procedure
NOTE
The screening mode must be enabled. Select
(Settings) > Device settings > Set up camera >
Screening mode = On.
The screening alarm deactivates any setting that puts the camera in standby mode.
Follow this procedure:
1. Push the navigation pad to display the menu system.
2. Use the navigation pad to go to
(Recording mode).
3. Push the navigation pad. This displays a submenu.
4. Select
(Screening).
5. Push the navigation pad. This displays a dialog box where you can define the settings
for the alarm.
Allowed deviation: The allowed deviation from the sampled average.
Alarm sound: Applicable values are Beep or No sound.
6. Push the navigation pad. This closes the dialog box.
7. Aim the camera toward a point of interest. The object should be within the frame of
the measurement box.
8. Tap and hold the
icon to reset the sampled average.
9. Tap the
icon to sample.
10. Aim the camera toward more points of interest. Sample 10 times to build up a sample
base by tapping the
icon.
The alarm is now set up and ready to use. Occasionally record a few samples if the
alarm is used for a long time or if the conditions change.
NOTE
The algorithm has a memory of the last 10 samples. It discriminates between the highest and lowest
values, and calculates an average of the remaining values.
Do not modify the measurement setup or activate another alarm because this will deactivate the
screening alarm.
#T559845; r. AJ/37554/37554; en-US
48

Table of Contents

Other manuals for FLIR E60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR E60 and is the answer not in the manual?

FLIR E60 Specifications

General IconGeneral
Resolution320 x 240 pixels
Thermal Sensitivity<0.05°C at 30°C
Field of View25° x 19°
Battery Life4 hours
Spatial Resolution1.36 mrad
Image Frequency60 Hz
MeterLinkYes
Laser PointerYes
Text AnnotationYes
Storage MediaSD card
Spectral Range7.5 – 13 µm
Temperature Range-20°C to +650°C
Display3.5" LCD touchscreen
FocusManual
Accuracy±2°C or ±2% of reading
Image ModesThermal, visual, MSX
Voice AnnotationYes
Battery TypeLi-Ion battery
Minimum Focus Distance0.4 m
Operating Temperature Range-15°C to +50°C
Storage Temperature Range-40°C to +70°C
ConnectivityWi-Fi, USB
Image StorageJPEG

Related product manuals