6 - 22 92B/96B/99B/105B Users Manual
ALTERING THE CONTINUOUS AUTO SET
CONFIGURATION
When is pressed, the instrument assumes the
Continuous Auto Set configuration.
There are ways to alter the Continuous Auto Set configura-
tion. These two methods are discussed below.
Using Manual Override
Pressing
disables autoranging for voltage attenua-
tion only. Press to resume the normal mode of opera-
tion for Continuous Auto Set.
Pressing disables autoranging for the time base
setting only. Press to resume the normal mode of
operation for Continuous Auto Set.
Changing any of the trigger settings disables autoranging for
voltage, time base, and triggering. Press to resume
the normal mode of operation for Continuous Auto Set.
Changing Continuous Auto Set Options
Change the Continuous Auto set configuration as follows:
1. From the User Options menu, press (MORE
OPTIONS) and use
to highlight AUTO SET.
Press to open the window (see Figure 6-14).
Figure 6-14. Continuous Auto Set Window