8 - 6 92B/96B/99B/105B Users Manual
6. Use to highlight 10:1, and press to
select.
7. Use to select CALIBRATE PROBE on A, and
press to calibrate.
The test tool outputs a voltage on the GENERATOR OUT
banana jack to the connected probe.
8. Adjust the trimmer screw in the probe housing of
INPUT A to give an optimum square wave within the
equal markers on the display (see Figure 8-3).
Figure 8-3. Adjusting the Probe
9. Press to continue.
The test tool automatically calibrates itself to the probe.
A message appears on the display indicating that
calibration is successfully completed.
10. Press to close the procedure.
INPUT B Probe Calibration
1. Connect the INPUT B BNC input to the GENERATOR
OUT red banana jack. See Figure 8-4. Use the grey
scope probe and the grey 4-mm banana adapter.
Figure 8-4. INPUT B Probe Calibration Setup
2. Repeat steps 2 trough 10 of the INPUT A calibration.