EasyManua.ls Logo

Fluke 117 - Accuracy Specifications - Voltage

Fluke 117
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
114, 115, and 117
Users Manual
20
Certifications............................................................UL, P, CSA, TÜV, ;
(N10140), VDE
IP Rating (dust and water protection) ......................... IP42
Table 1. Accuracy Specifications
Function Range Resolution
Accuracy
± ([% of Reading] + [Counts])
Model
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2
114,
115, 117
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2
114,
115, 117
DC, 45 to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ
[1]
True-rms 600.0 V 0.1 V
2.0 % + 3 4.0 % + 3
114, 117
45 to 500 Hz
500 Hz to 1 kHz
AC millivolts
[1]
True-
rms
600.0 mV 0.1 mV
1.0 % + 3
2.0 % + 3
114,
115, 117
AC Volts
[1]
True-rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3
2.0 % + 3
114,
115, 117