Characteristics
2.1 Introduction 2
2-3
2.1 Introduction
Performance Characteristics
FLUKE guarantees the properties expressed in numerical values with the stated tolerance.
Specified non-tolerance numerical values indicate those that could be nominally expected
from the mean of a range of identical ScopeMeter test tools.
Specifications are based on a one year calibration cycle.
Environmental Data
The environmental data mentioned in this manual are based on the results of the
manufacturer’s verification procedures.
Safety Characteristics
The test tool has been designed and tested in accordance with Standards ANSI/ISA
S82.02.01, EN 61010-1: 2001, CAN/CSA-C22.2 No.61010-1-04 (including
c
CSA
us
approval), Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use.
Use of this equipment in a manner not specified by the manufacturer may impair
protection provided by the equipment.
2.2 Dual Input Oscilloscope
2.2.1 Vertical
Frequency Response
DC Coupled:
excluding probes and test leads:
Fluke 123 (via BB120) DC to 20 MHz (-3 dB)
Fluke 124, 125 (via BB120) DC to 40 MHz (-3 dB)
with STL120 1:1 shielded test leads: DC to 12.5 MHz (-3 dB)
DC to 20 MHz (-6 dB)
with VP40 10:1 probe:
Fluke 123 (optional accessory) DC to 20 MHz (-3 dB)
Fluke 124, 125 (standard accessory) DC to 40 MHz (-3 dB)
AC Coupled (LF roll off):
excluding probes and test leads <10 Hz (-3 dB)
with STL120 <10 Hz (-3dB)
with 10:1 10MΩ Probe <1 Hz (-3 dB)
Rise Time
excluding probes and test leads:
Fluke 123 <17.5 ns
Fluke 124, 125 <8.75 ns
Input Impedance
excluding probes and test leads 1 MΩ//12 pF
with BB120 1 MΩ//20 pF
with STL120 1 MΩ//225 pF
with VP40 10:1 Probe 5 MΩ//15.5 pF