EasyManua.ls Logo

Fluke 123 - Environmental

Fluke 123
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 123/124/125
Service Manual
2-12
Memory
Number of Screen + Setup Memories
Fluke 123 10
Fluke 124 20
Number of Data Set memories
Fluke 125 20
Mechanical
Size 232 x 115 x 50 mm (9.1 x 4.5 x 2 in)
Weight 1.2 kg (2.5 lbs), including battery pack.
Interface RS-232, optically isolated
To Printer supports Epson FX, LQ, and HP Deskjet
®
,
Laserjet
®
, and Postscript
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional).
Parallel via PAC91 (optically isolated print
adapter cable, optional).
To PC Dump and load settings and data.
Serial via OC4USB (optically isolated
RS232/USB adapter/cable, optional), using
SW90W (FlukeView software for
Windows).
2.8 Environmental
Environmental MIL-PRF-28800F, Class 2
Temperature
Operating 0 to 50 °C (32 to 122 °F)
Storage -20 to 60 °C (-4 to 140 °F)
Humidity
Operating:
@0 to 10 °C (32 to 50 °F) noncondensing
@10 to 30 °C (50 to 86 °F) 95%
@30 to 40 °C (86 to 104 °F) 75%
@40 to 50 °C (104 to 122 °F) 45%
Storage:
@-20 to 60 °C (-4 to 140 °F) noncondensing
Altitude
Operating 5 km (16 400 feet)
Max. Input and Floating Voltage 600 Vrms
Cat III up to 2 km, > 2 km 300 Vrms Cat
III < 5 km
Storage 12 km (40 000 feet)
Vibration (Sinusoidal) MIL28800F, Class 2, 3.8.4.2, 4.5.5.3.1,
Max. 3g
Shock MIL28800F, Class 2, 3.8.5.1, 4.5.5.4.1,
Max. 30g
Fungus Resistance MIL28800F, Class 3, 3.8.7 & 4.5.6.1

Table of Contents

Other manuals for Fluke 123

Related product manuals