EasyManuals Logo

Fluke 123 Service Manual

Fluke 123
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Calibration Adjustment
4.6 Final Calibration 4
4-9
Table 4-2. HF Gain Calibration Points Slow
Cal step 5500A Setting
(1 kHz, MODE
wavegen,
WAVE square)
Test Tool Input Signal
Requirements
(1 kHz square, t
rise
<2 μs,
flatness after rising edge:
<0.5% after 4 μs)
HF-Gain AB (CL 0609) 25V 25V
HF-Gain A (CL 0612),
HF-Gain B (CL 0632)
HF-Gain A (CL 0615), HF-Gain B (CL 0635)]
1)
50V 50V
1)
After starting the first step in this table cell, these steps are done automatically.
4.6.2 Delta T Gain, Trigger Delay Time & Pulse Adjust Input A
Proceed as follows to do the calibrations:
1. Press
to select calibration step Delta T (CL 0700):IDLE
2. Connect the test tool to the 5500A as shown in Figure 4-4.
ST8004.WMF
Figure 4-4. 5500A Scope Output to Input A
3. Set the 5500A to source a 1V, 1 MHz fast rising (rise time 1 ns) square wave
(SCOPE output, MODE edge).
4. Set the 5500A to operate (OPR).
5. Press
to start the calibration.
The Delta T gain, Trigger Delay (CL0720), and Pulse Adjust Input A (CL0640) will
be calibrated.
6. Wait until the display shows Pulse Adj A (CL 0640):READY.
7. When you are finished, set the 5500A to Standby.
8. Continue at Section 4.6.3.

Table of Contents

Other manuals for Fluke 123

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 123 and is the answer not in the manual?

Fluke 123 Specifications

General IconGeneral
Bandwidth20 MHz
Sample Rate25 MS/s
Channels2
Input Impedance1 MΩ
Display Size5.7 inches
Temperature Range0 °C to 50 °C
Dimensions232 mm x 115 mm x 50 mm
DisplayDigital LCD
Memory Depth512 points per input
Battery Life7 hours
Weight1.2 kg
Safety CategoryCAT III 600 V

Related product manuals