Characteristics
2.3 Dual Input Meter 2
2-9
Open Circuit Voltage <4V
Continuity (CONT)
Beep 30Ω ± 5Ω in 50Ω range
Measurement Current 0.5 mA
Detection of shorts of 1 ms
Diode
Maximum Voltage:
@0.5 mA >2.8V
@open circuit <4V
Accuracy ±(2% +5 counts)
Measurement Current 0.5 mA
Polarity + on input A, - on COM
Capacitance (CAP)
Ranges 50 nF, 500 nF, 5 μF, 50 μF, 500 μF
Accuracy ±(2% +10 counts)
Full Scale Reading 5000 counts
Measurement Current 5 μA to 0.5 mA, increases with increasing
ranges
Measurement principle Dual slope integrating measurement with
parasitic serial and parallel resistance
cancellation.
2.3.3 Advanced Meter Functions
Zero Set Set actual value to reference
Fast/Normal/Smooth
Meter settling time Fast 1s @ 1μs to 10 ms/div
Meter settling time Normal 2s @ 1μs to 10 ms/div
Meter settling time Smooth 10s @ 1μs to 10 ms/div
Touch Hold (on A) Captures and freezes a stable measurement
result. Beeps when stable. Touch Hold
works on the main meter reading , with
threshholds of 1 Vpp for AC signals and
100mV for DC signals.
TrendPlot Graphs meter readings of the Min and Max
values from 15 s/div (120 seconds) to 2
days/div (16 days) with time and date
stamp. Automatic vertical scaling and time
compression.
Displays the actual and Minimum,
Maximum, or average (AVG) reading.
Fixed Decimal Point Possible by using attenuation keys.