Performance Verification
3.5 Input A and Input B Tests 3
3-9
9. Set the 5500A to source a 40 MHz (Fluke 123) or 60 MHz (Fluke 124/125) leveled sine
wave of 1.8V peak-to-peak.
10. Adjust the amplitude of the sine wave to 4 divisions on the test tool display.
11. Verify that the signal is well triggered.
If it is not, press
to enable the up/down arrow keys for Trigger Level adjustment;
adjust the trigger level and verify that the signal will be triggered now.
12. When you are finished, set the 5500A to Standby.
3.5.3 Input A Frequency Response Upper Transition Point Test
Proceed as follows to test the Input A frequency response upper transition point:
1. Connect the test tool to the 5500A as for the previous test (see Figure 3-3).
2. Select the following test tool setup:
Press
to select auto ranging (AUTO in top of display).
Do not press
anymore!
Using
change the sensitivity to select manual sensitivity ranging, and lock
the Input A sensitivity on 200 mV/div.
3. Set the 5500A to source a leveled sine wave of 1.2V peak-to-peak, 50 kHz (SCOPE
output, MODE levsine). Set the 5500A to Operate (OPR).
4. Adjust the amplitude of the sine wave to 6 divisions on the test tool display.
5. Set the 5500A to 20 MHz (Fluke 123) or 40 MHz (Fluke 124/125), without changing the
amplitude.
6. Observe the Input A trace check to see if it is 4.2 divisions.
7. When you are finished, set the 5500A to Standby.
Note
The lower transition point is tested in Section 3.5.11.
3.5.4 Input A Frequency Measurement Accuracy Test
Proceed as follows to test the Input A frequency measurement accuracy:
1. Connect the test tool to the 5500A as for the previous test (see Figure 3-3).
2. Select the following test tool setup:
Press
to select auto ranging (AUTO in top of display).
Press
to open the INPUT A MEASUREMENTS menu, and choose:
MEASURE on A: Hz
3. Set the 5500A to source a leveled sine wave of 600 mV peak-to-peak (SCOPE output,
MODE levsine). Set the 5500A to Operate (OPR).
4. Set the 5500A frequency according to the first test point in Table 3-1.