Performance Verification
3.5 Input A and Input B Tests 3
3-17
6. Continue through the test points.
7. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
Table 3-2. Volts DC Measurement Verification Points
Sensitivity
(Oscilloscope)
Range
1)
(Meter)
5500A output,
V DC
Input A-B DC Reading
5 mV/div 500 mV 15 mV 014.4 to 015.6
2)
10 mV/div 500 mV 30 mV 029.3 to 030.7
2)
20 mV/div 500 mV 60 mV 059.2 to 060.8
50 mV/div 500 mV 150 mV 148.7 to 151.3
100 mV/div 500 mV 300 mV 298.0 to 302.0
200 mV/div 500 mV 500 mV 497.0 to 503.0
-500 mV -497.0 to -503.0
0 mV -000.5 to + 000.5
500 mV/div 5V 1.5V 1.487 to 1.513
1 V/div 5V 3V 2.980 to 3.020
2 V/div 5V 5V 4.970 to 5.030
-5V -4.970 to -5.030
0V -0.005 to +0.005
5 V/div 50V 15V 14.87 to 15.13
10 V/div 50V 30V 29.80 to 30.20
20 V/div 50V 50V 49.70 to 50.30
-50V -49.70 to -50.30
0V -00.05 to +00.05
50 V/div 500V 150V 148.7 to 151.3
100 V/div 500V 300V 298.0 to 302.0
1)
The 500V and 1250V range will be tested in Section 3.5.14
2)
Due to calibrator noise, occasionally OL (overload) can be shown.