Performance Verification
3.5 Input A and Input B Tests 3
3-29
3. Set the TV Signal Generator to source a signal with the following properties:
the system selected in step 2
gray scale
video amplitude 1V (5 divisions on the test tool)
chroma amplitude zero.
4. Observe the trace, and check to see if the test tool triggers on line number:
622 for PAL or SECAM, see Figure 3-11
525 for NTSC, see Figure 3-12.
Note
Numerical readings in the pictures shown below may deviate from those shown
in the test tool display during verification.
PAL622.BMP
Figure 3-11. Test Tool Screen for PAL/SECAM
line 622
NTSC525.BMP
Figure 3-12. Test Tool Screen for NTSC line 525
5. Using select the line number:
310 for PAL or SECAM
262 for NTSC
6. Observe the trace, and check to see if the test tool triggers on:
line number 310 for PAL or SECAM, see Figure 3-13.
line number 262 for NTSC, see Figure 3-14.