150X Series
Calibration Manual
16
Required Tools and Equipment
Table 2 lists the required equipment for performing the Performance Test and Calibration
adjustments. If a recommended model is not available, use a substitute with equivalent or
better specifications
Table 2. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator AC Voltage Range: 0 - 600 V
Accuracy: +/- 0.5%
Frequency Range: 50 Hz - 400 Hz
Accuracy: +/-3%
DC Voltage: 0 - 600 V
Accuracy: +/-0.5%
Ohms Range: 0 - 20 kΩ
Accuracy: +/-0.37 %
Insulation Resistance: 50 kΩ - 60 MΩ
Accuracy: +/-0.37%
Fluke 9100 w/135 option
DMM DC Volts: 0 - 1000 V
Accuracy: +/-5%
DC Current: 0 - 2 mA
Accuracy: +/-1%
Fluke 189
HV Divider DC Voltage Range: 1 kV -5 kV
Accuracy: 1%
Division Ratio: 1000:1
Input resistance; 1000 MΩ
Fluke 80k-6
HV Resistor 9 GΩ Resistor for Insulation Test Ohmite # MOX-400239007F
Precision Resistor Resistor, wirewound, 2 Ω, +- 0.1%, 3 W,
50ppm, axial
IRC # RWR89S2R00BR
Precision Resistor Resistor, R05a, MF, 50 k, +-0.1%, 0.25 W,
25 ppm
Fluke PN 810580
Downloaded from Elcodis.com electronic components distributor