EasyManua.ls Logo

Fluke 150 Series - Required Tools and Equipment for Testing

Fluke 150 Series
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
150X Series
Calibration Manual
16
Required Tools and Equipment
Table 2 lists the required equipment for performing the Performance Test and Calibration
adjustments. If a recommended model is not available, use a substitute with equivalent or
better specifications
Table 2. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator AC Voltage Range: 0 - 600 V
Accuracy: +/- 0.5%
Frequency Range: 50 Hz - 400 Hz
Accuracy: +/-3%
DC Voltage: 0 - 600 V
Accuracy: +/-0.5%
Ohms Range: 0 - 20 k
Accuracy: +/-0.37 %
Insulation Resistance: 50 k - 60 M
Accuracy: +/-0.37%
Fluke 9100 w/135 option
DMM DC Volts: 0 - 1000 V
Accuracy: +/-5%
DC Current: 0 - 2 mA
Accuracy: +/-1%
Fluke 189
HV Divider DC Voltage Range: 1 kV -5 kV
Accuracy: 1%
Division Ratio: 1000:1
Input resistance; 1000 M
Fluke 80k-6
HV Resistor 9 G Resistor for Insulation Test Ohmite # MOX-400239007F
Precision Resistor Resistor, wirewound, 2 , +- 0.1%, 3 W,
50ppm, axial
IRC # RWR89S2R00BR
Precision Resistor Resistor, R05a, MF, 50 k, +-0.1%, 0.25 W,
25 ppm
Fluke PN 810580
Downloaded from Elcodis.com electronic components distributor