150X Series
Calibration Manual
30
Table 9. Calibration Adjustment Steps
Switch Position Input Terminal Adjustment Step Input Value
C-01 0m A, 0 Hz
C-02 15u A, 0 Hz
C-03 0.18 mA, 0 Hz
1000 V Insulation
C-04 1.8 mA, 0 Hz
C-05 0.5 mA, 0 Hz
C-06 5.0 mA, 0 Hz
C-07 0.5 mA, 0 Hz
C-08 5.0 mA, 0 Hz
C-09 5.0 mA, 0 Hz
Continuity
+ : COM
- : A gnd (remote test
probe pin)
C-10 300 mA, 0 Hz
C-11 25.0 V, 0 Hz
Volts
+ : Volts Input
- : COM
C-12 750 V, 0 Hz
C-13 0.5 V, 0 Hz
Continuity
+ : Continuity Input
- : COM
C-14 5.0 V, 0 Hz
1000 V Insulation None C-15
Continuity
+: Continuity Input
-: COM
C-16
2.00 (Use external
resistor, 0.1 %, 3 W,
50 PPM)
[1]
Any
+: Battery + Terminal
-: Battery - Terminal
C-18 + 5 V
[1] Must certify that this resistor is within 1.998 to 2.002 . This resistor should be mounted directly to the
Meter and COM terminals to minimize lead resistance.