Fluke 1550B - Electrical Specifications; Measurement Ranges and Accuracy

Fluke 1550B
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MegOhmMeter
Specifications
23
Electrical Specifications
The meter’s accuracy is specified for 1 year after calibration at operating
temperatures of 0 °C to 35 °C. For operating temperatures outside the range
(20 °C to 0 °C and 35 °C to 50 °C), add ± .25 % per °C, except on the 20 %
bands add ± 1 % per °C.
Insulation
Test Voltage (DC) Insulation Resistance Range Accuracy (+/- reading)
250 V <200 k
200 k to 5 G
5 G to 50 G
>50 G
unspecified
5 %
20 %
unspecified
500 V <200 k
200 k to 10 G
10 G to 100 G
>100 G
unspecified
5 %
20 %
unspecified
1000 V <200 k
200 k to 20 G
20 G to 200 G
>200 G
unspecified
5 %
20 %
unspecified
2500 V <200 k
200 k to 50 G
50 G to 500 G
>500 G
unspecified
5 %
20 %
unspecified
5000 V <200 k
200 k to 100 G
100 G to 1 T
>1 T
unspecified
5 %
20 %
unspecified
Bar graph range:
Insulation test voltage accuracy:
Induced ac mains current rejection:
Charging rate for capacitive load:
0 to 1 T
-0 %, +10 % at 1 mA load current
2 mA maximum
5 seconds per µF
Discharge rate for capacitive load 1.5 s/µF
Leakage current measurement
Range Accuracy
1 nA to 2 mA +/- (5 % + 2 nA)
Capacitive measurement
Range Accuracy
0.01 µF to 15.00 µF +/- (15 % of reading + 0.03 µF)
Timer
Range Resolution
0 to 99 minutes
Setting: 1 minute
Indication: 1 second
Live circuit warning Warning range: 30 V to 660 V ac/dc, 50/60 Hz
Voltage accuracy: +/- (5 % + 2 V)