Calibration Adjustment – Standalone Mode
You can do calibration adjustment in standalone mode only through the remote interface.
See Chapter 3 for command instructions. It is recommended that you do the verification
procedure above to find out if adjustments are necessary. Do a full verification after
adjustments are made.
Equipment Set Up
To set up the test equipment for calibration adjustments:
1. Open the link between V-GUARD and Ground terminals on the 5720A.
2. Set Remote Guard on the 5720A.
3. Connect the equipment as shown in Figure 5-2.