EasyManua.ls Logo

Fluke 52120A - Phase Angle Accuracy Confidence Check; Verify Amplitudes at DC and Higher Frequencies

Fluke 52120A
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
Introduction 5
5-23
Table 5-9. Power Line Frequency Confidence Check Points
6105A Measured Accuracy
Range RMS
Phase
Angle ()
V I I + Limit I - Limit
1 x Ext 2A 2 A 60° 2.000280 1.99720
1 x Ext 20A 20 A 60° 20.00280 19.99720
1 x Ext 120A 120 A 60° 120.0168 119.9832
Phase Angle Accuracy Confidence Check
To do a phase angle accuracy confidence check, set the range, rms, and phase angle of
the 6105A to the values in Table 5-10.
Table 5-10. Phase Angle Accuracy Confidence Check Points
6105A Measured Accuracy
Range RMS
Phase
Angle ()
Phase Angle
Error ()
PF VA Watts
1 x Ext 2A 2 A 60°
1 x Ext 20A 20 A 60°
1 x Ext 120A 120 A 60°
To calculate phase angle from PF or VA and watts:
PF =
Watts
VA
error() = cos
–1
(PF) –
θ
For example, if the power meter shows PF = 0.499894:
error() = cos
–1
(0.499894) – 60 = 0.007°
Verify Amplitudes at DC and Higher Frequencies
Use the Fluke A40B and 8508A to verify amplitude accuracy.
2 Amp Range
1. Set the DMM to DCV function on the 2 V range.
2. Set the DMM ‘m’ store to the 2 A shunt correction factor.
3. Set the DMM ‘-c’ and ‘÷z’ stores to 0.4.
4. Turn on the ‘*m’, ‘-c’ and ‘÷z’, and % math functions on the DMM.
5. Set the 6105A range to 1 x Ext 2A.
6. Set the current and frequency of the 6105A to the values in Table 5-11 for the 1 A
verification point.
7. Set the DMM ‘-c’ and ‘÷z’ stores to 0.8.
8. Set the DMM to ACV function.
9. Set the current and frequency of the 6105A to the values in Table 5-11 for the two
2 A verification point.

Table of Contents

Related product manuals