EasyManua.ls Logo

Fluke 52120A - Page 94

Fluke 52120A
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52120A
Users Manual
5-24
10. Set the DMM ‘-c’ and ‘÷z’ stores to 0.24.
11. Set the current and frequency of the 6105A to the values in Table 5-11 for the two
0.6 A verification points.
12. Set the DMM to the 200 mV range.
13. Set the DMM ‘-c’ and ‘÷z’ stores to 0.16.
14. Set the current and frequency of the 6105A to the values in Table 5-11 for the 0.4 A
verification point.
Table 5-11. Amplitude Verification on 2 Amp Range
6105A
Measured
Error (%)
Product
Specification
Measurement
Uncertainty
TUR
RMS Frequency
6105A
(99 %)
6100B
(99 %)
6105A 6100B
1 A DC ±0.0350 % ±0.0720 % ±0.0029 % 12:1 25:1
2 A 57 Hz ±0.0170 % ±0.0210 % ±0.0125 % 1.1:1 1.7:1
2 A 850 Hz ±0.0170 % ±0.0210 % ±0.0106 % 1.3:1 1.9:1
0.6 A
[1]
1083 Hz ±0.0687 % ±0.0687 % ±0.0175 % 3.9:1 3.9:1
0.6 A
[2]
5643 Hz ±0.0687 % ±0.0687 % ±0.0175 % 3.9:1 3.9:1
0.4 A
[3]
5643 Hz ±0.0770 % ±0.0770 % ±0.0154 % 5.0:1 5.0:1
[1] Fundamental 57 Hz, zero amplitude, 19
th
harmonic at 0.6 A
[2] Fundamental 57 Hz, zero amplitude, 50
th
harmonic at 0.6 A
[3] Fundamental 57 Hz, zero amplitude, 99
th
harmonic at 0.4 A
20 Amp Range
1. Set the DMM to the DCV function and the 2 V range.
2. Set the DMM ‘m store to the 20 A shunt correction factor.
3. Set the DMM ‘-c’ and ‘÷z’ stores to 0.4.
4. Turn on the ‘*m’, ‘-c’ and ‘÷z’, and % math functions on the DMM.
5. Set the 6105A range to 1 x Ext 20A.
6. Set the current and frequency of the 6105A to the values in Table 5-12 for the 10 A
verification point.
7. Set the DMM ‘-c’ and ‘÷z’ stores to 0.8.
8. Set the DMM to ACV function
9. Set the current and frequency of the 6105A to the values in Table 5-12 for the two
20 A verification points.
10. Set the DMM ‘-c’ and ‘÷z’ stores to 0.24.
11. Set the current and frequency of the 6105A to the values in Table 5-12 for the two
6 A verification points.
12. Set the DMM to the 200 mV range.
13. Set the DMM ‘-c’ and ‘÷z’ stores to 0.16.
14. Set the current and frequency of the 6105A to the values in Table 5-12 for the 4 A
verification point.

Table of Contents

Related product manuals