5790B
Operators Manual
6-38
CAL_STORE?
Tells how many calibration constant groups have been updated but not saved.
(This is like CAL_STLST? except only the initial number is given.)
Parameters
DC (DC calibration)
AC (AC calibration)
ZC (DC zeros calibration)
WDC (Wideband 1 kHz calibration)
WAC (Wideband Flatness)
ALL (All of the above; use this to check CAL_I2 constants)
Response
(Integer) 0 if no calibration constant groups have been updated; otherwise,
the number of groups.
CAL_SW?
Returns the calibration security state (secure or unsecure). (CAL_MODE?
returns the CALIBRATION MODE state setting).
Parameter
None
Response
(Integer) 0 = NORMAL (secure), 1 = unsecure
CAL_TEMP
Sets the temperature for calibration. Do this before you run any calibration
procedures. Once set, the temperature is used for all calibration activities until
changed. If not set before a calibration activity is complete, the Product uses a
default of 23.0 °C.
Parameter
A number signifying the temperature in °C.
CAL_TEMP?
Returns the temperature used for all calibration activities (ie: the value set by
CAL_TEMP).
Parameter
None
Response
(Real) The temperature in °C.