4. Touch the parameter you want to change. See Table 5-2.
5. Touch EOL to set the EOL (End Of Line) character to CR, LF, or the string
CR LF.
6. To set up the interface for remote control of the Product, touch Remote IF
and make a selection (Terminal or Computer).
Terminal sets the remote port to expect a human operator to be using a
terminal attached to the RS-232 port to control the Product. This setting
has the same effect as specifying TERM in the SP_SET or REM_MODE
remote command parameter string.
Computer sets the remote port to expect a computer to be controlling the
Product over the RS-232 port. This setting has the same effect as
specifying “COMP” in the SP_SET or REM_MODE remote command
parameter string.
7. To designate a character or character string as the EOF (End Of File)
character, touch Set EOF.
a. Enter the decimal code of the ASCII character(s) designated as EOF.
(Appendix A contains a table of ASCII codes.)
b. Verify that the selection is correct as shown on the display.
8. To exit the setup menus, touch Exit.
Serial Remote Control Setup Procedure
1. Turn on the Product.
2. Touch Setup Menu>System Settings>Remote Port.
3. If the Active Remote Port is not RS-232, touch Active Remote Port>RS-232.
4. Touch Exit to exit the Remote Port Menu.
Exceptions for Serial Remote Control
When you use the RS-232-C port to remotely control the Product, either
interactively with a terminal or under computer control, operation is the same as
when you use an IEEE-488 controller connected to the IEEE-488 port for control,
with the following exceptions:
Control/C does the same function as DCL (Device Clear) or SDC (Selected
Device Clear).
The EOL (End of Line) input terminator is Carriage Return (Control/M) or Line
Feed (Control/J). All output lines are terminated by the character selected in a
setup menu, or set using the remote commands SP_SET or EOL. This
setting applies to all lines, including those with the *PUD command (see
*PUD (Protected User Data) and RPTSTR).
Control/R echoes to the port a Carriage Return, a Line Feed, and any
uncompleted remote command entered. This allows you to see a copy of
whatever has been typed in since the last command.