2-1
Chapter 2
Installation
Warning
To avoid electric shock and to conform to IEC Safety Class I,
use the Product only with sources up to 1000 V dc or rms ac
that are protected from short circuit with current-limiting to
200 mA or less.
Introduction
This chapter provides instructions for installing the Product and connecting it to
line power. Because this chapter explains fusing and operating environment
requirements, you should read this chapter before operating the instrument.
Instructions for connecting cables to other standards and to a UUT (Unit Under
Test) during operation are in Chapter 4.
Unpacking and Inspection
The Product ships in a container that prevents shipping damage. Inspect the
Product carefully for damage, and immediately report any damage to the shipper.
Instructions for inspection and claims are included in the shipping container.
When the Product is unpacked, check for all the standard equipment listed in
Table 2-1 and check the shipping order for additional items ordered. Report any
shortage to the place of purchase or to the nearest Fluke Calibration Service
Center. See How to Contact Fluke Calibration if necessary. If performance tests
are required for your acceptance procedures, refer to the Product service manual
for instructions.