Remote Operation
RS-232 Serial Interface 5
5-5
Set Up and Connect the Serial Interface
Refer to the specifications for the peripheral device, and proceed as follows to
set up the serial interface for the application:
1. With the Product power off, connect a shielded 9-pin D subminiature RS-232
null-modem cable such as Fluke accessory RS43 to the rear panel RS-232
connector and to the peripheral device. Always use a completely shielded
cable.
2. Push to turn on the Product.
3. Touch Setup Menu>System Settings>Remote Port>RS-232 Setup. The
display changes to:
System
Settings
Remote
Port
RS-232
Setup
Setup Menu
Data Bits 8
Stop Bits 1
Stall Xon/Xoff Remote IF Terminal
EOL CRLF
Parity None
Baud 9600
Set EOF 000 000
Restore Factory Defaults
2/05/15 10:52am Setup
Input 2: 7.54-- mV
Exit
hvi022.eps