5790B
Operators Manual
6-42
DELTA?
Returns the delta (transfer) value from the Product. This is the difference
between the reference value in memory and the present input. (Also see the
DUNIT command.)
Parameters
None
Responses:
1. (Float) the transfer value. The value is 0 if no reference has been set.
2. (String) PPM, PCT, V, RATIO as set by the DUNIT command.
DFILT
Sets the digital filter parameters. The default is OFF.
Parameter
1. OFF, SLOW, MEDIUM, or FAST (the setting for mode)
2. (Optional) FINE, COARSE, or MEDIUM (the setting for restart threshold)
DFILT?
Returns the digital filter parameter settings.
Parameter
None
Response
1. (CRD) OFF, SLOW, MEDIUM, or FAST (the setting for mode)
2. (CRD) FINE, COARSE, or MEDIUM (the setting for restart threshold)
DHCP
Sequential command. Enables/disables DHCP (Dynamic Host Configuration
Protocol) for LAN operation.
Parameter
ON (enables DHCP operation)
OFF (disables DHCP operation)
DHCP?
Sequential command. Returns the current state of the DHCP enable setting.
Parameter
None
Response
(CRD)
Example
DHCP?
Returns ON if DHCP is enabled.