Remote Commands
Alphabetical List of 5790B Device-Dependent Commands 6
6-51
number with #b, #o, or #h, respectively. (A number without a preceding
designator is taken as decimal.)
Example
ISCE1 52
Loads 00000000 00110100 (binary) into the register, enabling bits 5
(MCCHG), 4 (MDCHG), and 2 (RNGCHG).
ISCE1?
Reads the contents of the Instrument Status 0 to 1 Change Enable register
described under "Checking Product Status."
Parameter
None
Response
(Integer) The contents of the Instrument Status 0 to 1 Change Enable register
in decimal.
Example
ISCE1?
Returns 52 if bits 5 (MCCHG), 4 (MDCHG), and 2 (RNGCHG) are enabled (set
to 1) in the Instrument Status 0 to 1 Change Enable register.
ISCR0?
Reads and clears the contents of the Instrument Status 1 to 0 Change Register
described under "Checking Product Status."
Parameter
None
Response
(Integer) The contents of the Instrument Status 1 to 0 Change Register in
decimal.
Example
ISCR0?
Returns 52 if bits 5 (MCCHG), 4 (MDCHG), and 2 (RNGCHG) have been set
to 1 in the Instrument Status 1 to 0 Change register, and clears the register.
ISCR1?
Reads and clears the contents of the Instrument Status 0 to 1 Change Register
described under "Checking Product Status."
Parameter
None
Response
(Integer) The contents of the Instrument Status 0 to 1 Change Register in
decimal.
Example
ISCR1?
Returns 52 if bits 5 (MCCHG), 4 (MDCHG), and 2 (RNGCHG) have been set
to 1 in the Instrument Status 0 to 1 Change register, and clears the register.