Operators Manual
6-54
MW?
Returns (50 Ω) milliwatt power equivalent of the reading. This is valid only for
wideband.
Parameter
None
Response
1. (real) Amplitude
2. (String) Units
Example
MW?
1.2736105E+00,MW
ONTIME?
Returns the time in minutes that the Product power has been turned on for the
current operating session.
Parameters
None
Response
(Integer) The number of minutes.
PKPK?
Returns the peak-to-peak power equivalent for the requested waveform (SINE,
SQUARE, TRI, TRUNCS).
Parameter
SINE, SQUARE,TRI, or TRUNC
Response
1. (real) Amplitude
2. (String) Units
Example
PKPK? SQUARE
2.0000560E+00,V
RANGE
Selects the range that best measures the specified value furnished as a
parameter, or otherwise controls the range setting as defined by the parameter.
The range selected will be the one within whose limits (as returned by the
RANGE? command) the value falls. This command turns off autoranging if it was
on, unless the parameter is AUTO.
Parameter
The value to be measured (the V or mV unit is optional).
or:
LOCK (To lock the range)
AUTO (To select autoranging)
UP (To go to the next highest range)
DOWN (To go to the next lowest range)