Remote Commands
Alphabetical List of 5790B Device-Dependent Commands 6
6-63
SH_LECLR
Clears all of the loading error corrections for the currently selected shunt.
Parameter
None
SH_LEDEL
Deletes a loading error correction for the currently selected shunt.
Parameter
1. (Integer) Numeric index of the loading error correction (where zero
represents the first element in the table)
SH_LEDEL
SH_ACDEL 2
Deletes the 3rd value from the loading error correction table
SH_LEMOD
Sets a loading error correction for the currently-selected shunt to a new
frequency and error.
Parameter
1. (Integer) Numeric index of the loading error correction (where zero
represents the first element in the table)
2. (Real) Frequency
3. (Integer) Error in ppm
Example
SH_LEMOD 2,2e4,5
Sets the 3rd value of the loading error correction table to 20.0 kHz and 5 ppm
SH_LESETALL (variable)
Sets all the loading error corrections at once.
The syntax is: SH_LESETALL numpts, freq in Hz,error in ppm,...
Parameter
1. (Integer) Number of data points
2. (Real) Frequency 1 in Hz
3. (Integer) Error 1 in ppm
.
.
.
Example
SH_LESETALL 3, 1.0e+3, -12, 5.5e+4, 8, 2.0e+5, -3