7102
User’s Guide
6
W Warning
To prevent personal injury:
• Be aware that the Product operates at high temperatures.
• Keep all flammable and meltable materials away from the Product.
• Although the Product is well insulated, top surfaces become hot.
• If the bath is operated at high temperatures, use a fume hood to remove any vapors
given off by hot bath fluid.
Bath Fluids (Summary)
The Product does not come with fluid. Various fluids are available from Fluke Calibration and other
sources. To safely operate the Product, only select bath fluids with adequate thermal properties that
meet the application requirements.
Depending on the desired temperature range, any of these fluids, as well as others, can be used:
• Ethylene glycol
• Mineral oil
• Silicone oil
See Bath Fluid and its subsections for information on fluid selection. See the specific MSDS sheet for
each fluid selected.
Installation
W
Caution
• Read Bath Use before you place the bath into service. incorrect handling can damage
the Product and void the warranty.
• Do not operate this Product without fluid.
• Make sure that the tank is 3/4 full before you set the temperature.
To install the Product:
1. Remove the access lid from the bath and check the tank for foreign matter (for example, dirt or
remnant packing material). Before you fill the well, thoroughly dry the inside of the well with paper
towels.
2. Plug the power cord into a grounded mains outlet. Observe that the nominal voltage corresponds to
that indicated on the back of the Product.
3. Carefully insert the probe basket into the well, then insert the probe(s) into the basket.
4. With the probe(s) in the well, fill the tank 3/4 full with the appropriate, clean fluid.