Micro-Bath
Scan
23
Temperature Scale Units
The selected units (Celsius (°C) or Fahrenheit (°F)) show in the displayed well temperature, set-point,
and proportional band.
Push SET after you adjust the set-point value to change display units.
Un= C Scale units currently selected
Push UP or DOWN to change the units.
Un= F New units selected
Scan
Set and enable the scan rate so that when the set-point is changed the bath heats or cools at a
specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the bath
heats or cools at the maximum possible rate.
Scan Control
Control the scan with the scan on/off function that appears in the main menu after the set-point
function.
Sc=OFF Scan function off
Push UP or DOWN to toggle the scan on or off.
Sc=On Scan function on
Push SET to accept the present setting and continue.