Micro-Bath
Well Temperature
21
Well Temperature
The display allows direct viewing of the actual well temperature. This temperature value is what is
normally shown on the display. The temperature units are displayed at the right. For example,
100.00C Well temperature in degrees Celsius
Push Exit to access the temperature display function from any function.
Temperature Set-point
Set the temperature set-point to any value within the range and resolution as given in the
specifications. Do not exceed the safe upper temperature limit of any device inserted into the well.
To set the temperature, select the set-point memory and adjust the set- point value.
Programmable Set-points
The controller stores eight set-point temperatures in memory. Quickly recall the set-points to
conveniently set the calibrator to a previously-programmed temperature set-point.
To set the temperature, select the set-point memory first. Push SET to access this function from the
temperature display function. The number of the current set-point memory shows at the left on the
display followed by the current set-point value.
100.00C Well temperature in degrees Celsius
Push SET Access set-point memory
1 25. Set-point memory 1. 25 °C currently used
To change the set-point memory push UP or DOWN.
4 125. New set-point memory 4 125. °C
Push SET to accept the new selection and access the set-point value.
Set-point Value
To adjust the set-point after you select the set-point memory, push SET.
4 125. Set-point 4 value in °C
If the set-point value is correct, hold EXIT to show the well temperature. Push UP or DOWN to adjust
the set-point value.
125.00 New set-point value
When the controller reaches the desired set-point value, push SET to accept the new value and to
access the temperature scale units selection. If you push EXIT instead of SET, any changes made to
the set-point are ignored.
Push SET Accept new set-point value