EasyManua.ls Logo

Fluke 725Ex - Page 33

Fluke 725Ex
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using Measure Mode
23
Table 5. Thermocouple Types Accepted
Positive Lead (H)
Color
Type
Positive Lead
Material
ANSI* IEC**
Negative Lead
Material
Specified Range
(°C)
E Chromel Purple Violet Constantan -200 to 950
N Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300
J Iron White Black Constantan -200 to 1200
K Chromel Yellow Green Alumel -200 to 1370
T Copper Blue Brown Constantan -200 to 400
B Platinum (30 % Rhodium) Gray Platinum (6 % Rhodium) 600 to 1800
R Platinum (13 % Rhodium) Black Orange Platinum -20 to 1750
S Platinum (10 % Rhodium) Black Orange Platinum -20 to 1750
L Iron Constantan -200 to 900
U Copper Constantan -200 to 400
GOST
XK
90.5 % Ni + 9.5 % Cr
Violet or Black
56 % Cu + 44 % Ni
-200 to 800
BP 95 % W + 5 % Re Red or Pink 80 % W + 20 % Re 0 to 2500
*American National Standards Institute (ANSI) device negative lead (L) is always red.
**International Electrotechnical Commission (IEC) device negative lead (L) is always white.

Table of Contents

Other manuals for Fluke 725Ex

Related product manuals