EasyManua.ls Logo

Fluke 725Ex - Calibrating a Pressure Transmitter

Fluke 725Ex
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating a Pressure Transmitter
43
Calibrating a Pressure Transmitter
The following example shows how to calibrate a pressure
transmitter.
Connect the Calibrator to the instrument under test as
shown in Figure 20. Proceed as follows:
1. Press
l for current (upper display). If required,
press l again to activate loop power.
2. Press
U (lower display).
3. If necessary, press
M for SOURCE mode.
4. Zero the pressure module.
5. Perform checks at 0 % and 100 % of span and adjust
the transmitter as necessary.

Table of Contents

Other manuals for Fluke 725Ex

Related product manuals