EasyManua.ls Logo

Fluke 725Ex - Calibrating an I;P Device

Fluke 725Ex
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating an I/P Device
45
Calibrating an I/P Device
The following test allows calibration of a device that
controls pressure. Proceed as follows:
1. Connect the test leads to the instrument under test as
shown in Figure 21. The connections simulate a
current-to-pressure transmitter and measures the
corresponding output pressure.
2. Press
A (upper display).
3. Press
V for sourcing current (lower display).
4. If necessary, press
M for SOURCE mode.
5. Enter the desired current by pressing
X and W
keys. Press Y and Z to select different digits.

Table of Contents

Other manuals for Fluke 725Ex

Related product manuals