725Ex
Users Manual
40
Auto Ramping the Output
Auto ramping gives the ability to continuously apply a
varying stimulus from the Calibrator to a transmitter, while
your hands remain free to test the response of the
transmitter.
When
L is pressed, the Calibrator produces a
continuously repeating 0 % - 100 % - 0 % ramp in a choice
of three ramp waveforms:
•
E 0 % - 100 % - 0 % 40-second smooth ramp
• P 0 % - 100 % - 0 % 15-second smooth ramp
• N 0 % - 100 % - 0 % Stair-step ramp in 25 %
steps, pausing 5 seconds at each step. Steps are
listed in Table 7.
To exit ramping, press any button.
Table 7. mA Step Values
Step 4 to 20 mA
0 % 4.000
25 % 8.000
50 % 12.000
75 % 16.000
100 % 20.000
Storing and Recalling Setups
Store up to eight settings in a nonvolatile memory and
recall the settings for later use. A low battery condition or a
battery change does not jeopardize the stored settings.
Proceed as follows:
1. After creating a calibrator setup, press
S. In the
display, the memory locations appear.
2. Press
Y or Z to select locations one through eight. An
underscore appears below the selected memory
location.
3. Press
S until the memory number disappears then
reappears. The setup is stored.
To recall setups, proceed as follows.
1. Press
Q. The memory locations appear on the
display.
2. Press
Y or Z to select the appropriate location and
press Q.