Making Measurements
Using the Measurement Functions
3
OHMS Resolution
Press the Resl key to see the OHMS RESL menu:
OHM RESL Menu
adj075f.eps
5
6
7
8
5-1/2-digits resolution
6-1/2-digits resolution
7-1/2-digits resolution
8-1/2-digits resolution
Refer to Table 3-2 for the ADC integration times applicable to the various resolution and
Fast On/Off settings.
To revert to the
OHMS CONFIG menu, press the Config key.
Ohms - Movement between Menus
CONFIG
TOGGLE
ON/OFF
TOGGLE
TOGGLE
adj076f.eps
Ohms Movement Between Menu
High Voltage Ohms
Press the PLUS key to see the expanded Ohms selection for True Ohms function and
the High Voltage Ohms function. Press
HiV key to see the HiV menu:
adj077f.eps
adj078f.eps
HiVMenu
The function provides both 2-wire and 4-wire measurements of resistance, in decade
ranges from 20 Mto 20 G
. The measurement is performed at High Voltage
using a current source with high compliance. The resulting increase in Current
through the unknown resistor is valuable to the reduction of leakage and bias
current uncertainties. The function may also be used alongside the normal Ohms
function to determine Voltage Coefficient in the unknown resistor.
3-13