4
Remote Operations Using the IEEE 488 Interface
Syntax Diagrams of Major 8508A Functions
Recall of Spot Frequency Value
SPOT? 1
2
3
4
5
6
phs
adj211f.eps
Recall of Spot Frequency
(Nr3) Recalls the frequency value for the specified calibrated spot on the active ACV
range.
Resistance: Normal OHMS
The following commands are used to select the Normal
OHMS function and set its
configuration.
,
OHMS
phs
Nrf
AUTO
FILT_ON
FILT_OFF
RESL5
RESL6
RESL7
RESL8
FAST_ON
FAST_OFF
TWO_WR
FOUR_WR
LOI_ON
LOI_OFF
adj212f.eps
Normal OHMS
Nrf
A numeric value that selects the range for the expected signal. Any valid numeric value
cancels autorange.
AUTO
Selects the autorange mode where the multimeter will determine the appropriate range for
the measured signal. If the signal exceeds the maximum capability then
RANGE
OVERLOAD
appears on the front panel and the appropriate bit is set in the device status
registers.
FILT_ON / FILT_OFF
FILT_ON
inserts a hardware analog filter into the measurement signal path.
4-13