Specifications
Specifications
5
Specifications
General Specifications
Power
Voltage
115 V Setting ..................................................... 100 V to 120 V rms
designed for additional voltage fluctuations ±10 %.
230 V Setting ..................................................... 200 V to 240 V rms
designed for additional voltage fluctuations ±10 %.
Frequency ............................................................... 50 Hz to 60 Hz designed for additional frequency variation of
±3 Hz.
Consumption ........................................................... < 80 VA
Power Cord ............................................................. NEMA 5-15 plug, IEC 60320-C13 receptacle, cable 3 core 18AWG to
SVT
Dimensions
Height ...................................................................... 88 mm (3.5 inches)
Width ....................................................................... 427 mm (16.8 inches)
Depth ...................................................................... 487 mm (19.2 inches)
Weight ..................................................................... 11.5 kg (25.5 lbs)
Environment
Temperature
Operating ............................................................ 0 °C to 50 °C
Specified Operation ............................................ 5 °C to 40 °C
Calibration (TCal)................................................ 20 °C to 25 °C
Factory Cal Temp ............................................... 23 °C
Storage ............................................................... -20 °C to 70 °C
Warm Up............................................................. 4 hours to full uncertainty specification.
Relative Humidity (non- condensing)
Operating
[18]
........................................................ 5 °C to 40 °C < 90 %.
Storage .............................................................. 0 °C to 70 °C < 95 %
Altitude
Operating ............................................................ < 2000 meters
Storage ............................................................... < 12000 meters
Vibration and Shock ................................................ Complies with MIL-PRF-28800F Class 3.
Safety ..................................................................... Designed and tested to IEC/EN61010-1: 2001,
UL 61010-1:2004,
CAN/CSA-C22.2 No.61010.1-04,
CE marked and ETL (US & C) listed.
Pollution Degree 2.
Installation Category II.
Equipment Class I (single insulation / Earthed metal case).
Protection against water ingress IP20 (general indoor conditions).
EMC ........................................................................ EN61326-1:2006 class B,
FCC Rules part 15 sub part B,
Measurement Isolation
Guard to Safety Ground ......................................... < 3300 pF, > 10 G.
Lo to Guard
In Remote Guard ................................................ < 2800 pF, > 10 G(Not in Resistance function).
In Local Guard .................................................... Lo and Guard terminals are internally shorted
(in Resistance < 2800 pF, > 10 G).
Autorange
Range Up................................................................ 100 % of range.
Range Down ........................................................... 9 % of range (18 % on 1000 V range).
Remote Interface ................................................... IEEE 488.2
Warranty ................................................................ 1 Year