8508A
Users Manual
Clear Status
CLS
*
adj265f.eps
Clear Status
*CLS
Clears all the event registers and queues except the output queue. The output queue and
MAV bit will be cleared if *CLS immediately follows a '"Program Message
Terminator".
Power On Status Clear
Nrf
phs
PSC
*
adj264f.eps
Power On Status Clear
*PSC
Sets the flag controlling the clearing of defined registers at Power On.
Nrf
A number which, when rounded to an integer value of zero, sets the power on clear flag
false. This allows the instrument to assert SRQ at power on.
When the value rounds to an integer value other than zero it sets the power on clear flag
true, which clears the standard event status enable and service request enable registers so
that the instrument will not assert an SRQ on power up. This data is saved in non volatile
memory for use at Power On.
Recall Status Clear Flag
PSC?
*
adj267f.eps
Recall Power On Status Clear)
*PSC?
(Nr1) Recalls the saved Power On status condition: false (0) and true (1).
Instrument I/D and Setup
I/D (Identification)
IDN?
*
adj269f.eps
I/D
∗IDN?
Recalls the instrument’s manufacturer, model number, serial number and firmware level.
Response
The data contained in the four comma separated fields is organized as follows:
• First field - Manufacturer
• Second field Model
• Third field - Serial number
4-44