8508A
Users Manual
Normal
Checks
1. Press the
key.
2. Press the
CONFIG key and select 4w. Ensure LoI is deselected.
3. Select the required 8508A range.
4. On the Calibrator, select zero and Operate. Alternatively if using a standard
resistor connect the standard resistor for a 4-wire resistance zero. (For the 200 M
range and above, de-select Ex SNS on the calibrator and de-select
4W οn the
8508A.)
5. On the 8508A press the
INPUT key, and select Zero Rng to initiate the zeroing
operation.
The
Busy legend will appear, and then extinguish when the zero operation has
completed.
6. On the Calibrator, select Range Gain point Output value. Alternatively if using a
standard resistor re-connect the standard resistor to measure its resistance.
7. Note the 8508A reading. Compare the result with the allowable tolerance, taking into
account the actual value of the calibrator resistance output for this value or the
calibrated value of the resistance standard used.
8. Repeat steps 3 – 7 for the remaining ranges of the selected resistance mode. (For the
2G and 20 G ranges the 8508-7000K 1 G standard resistor is used in place of
the calibrator. For the 200 M range and above, de-select Ex SNS on the calibrator
and de-select
4W οn the 8508A).
Normal LoI
checks
Follow the procedure for Normal checks ensuring that LoI is selected.
Press the key. Press the CONFIG key and select LoI.
Tru
checks
Follow the procedure for Normal checks except select the Plus key followed by Tru
instead of selecting the key (2 to 20 k ranges only).
Tru
LoI checks
Follow the procedure for Normal checks except select the Plus key followed by
Tru instead of selecting the key. Press the CONFIG key and select LoI (2 to 20 k
ranges only).
HiV
checks
Follow the procedure for Normal checks except select the Plus key followed by HiV
instead of selecting the key.
6-44