EasyManua.ls Logo

Fluke 8508A - Page 191

Fluke 8508A
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
Adjustment and Verification
Performance Verification
Equipment Configuration
1. On the 8508A, select
CLEAR, Pwr Up Dflt to select DCV 1 kV range.
2. Ensure that the calibrator output is set to
STANDBY and configured for Internal
Guard (
EX GRD is deselected).
3. Connect the calibrator to the 8508A as described for Routine Adjustment.
Note
After making the connections allow at least 30 minutes for thermal emfs in
the leads to stabilise.
4. On the calibrator select zero and
OPR to operate the output.
Select
EX SNS.
5. Confirm the 8508A is configured for internal guard (External Guard is deselected) by
checking that the
ExGd legend does not appear on the left hand display.
Procedure
Perform the following sequence for each resistance mode and each range, as listed in
Table 6-10, starting with the 2 range and working up to the 2 G range for the
Normal and Normal LoI resistance modes. Then continue the sequence for the Tru
and Tru LoI modes, starting with the 2 range and working up to the 20 k range.
Finally complete the sequence for the HiVmode, starting with the 20 M range and
working up to the 20 G range. An input zero operation is performed on each 8508A
range with the calibrator set to zero prior to performing the range verification
measurements. A separate zero operation is required on each range for each resistance
mode.
Note that EX SNS (4-wire) is not available on the 5720A 100 M range, and the 8508-
7000K 1 Gstandard resistor is used as a 2-wire device with the calibrator 2-wire
resistance zero used for the corresponding zero point. Ensure
4Wis deselected on the
8508A when checking the 200 M and 2 G range, range zero, and range gain points.
 Warning
The Multimeter can deliver a lethal electric shock when operating in the
High Voltage Resistance mode. Voltages up to 240V may appear on the
terminals.
To avoid electric shock when performing the following adjustment
procedures:
Never touch any lead or terminal unless you are absolutely certain
that no dangerous voltage is present.
Make sure that signal leads are in a safe condition before you
handle them in any way.
6-43

Table of Contents

Other manuals for Fluke 8508A

Related product manuals