EasyManua.ls Logo

Fluke 8508A - Ω Guard

Fluke 8508A
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
Making Measurements
Using the Measurement Functions
SENSE HI
SENSE HI
SENSE LO
SENSE LO
GUARD
GUARD
INPUT HI
INPUT HI
INPUT LO
INPUT LO
Rx
Metal
Screen
adj094f.eps
Figure 3-8. 4-Wire Resistance Zero Measurements
Guard
In the resistance function with Ext Grd selected the Guard terminal functions as
Guard. Using the Guard
terminal as Guard, the Guard feature can be used to make
‘in-circuit’ resistance measurements by guarding out parallel resistance paths so that only
the value of Rx will be displayed.
Similarly,
Guard can be used to reduce the settling time if Rx is shunted by any
capacitance and a suitable tapping point is available. The connections for making
Guard measurements are shown in Figure 3-9. Selection of External Guard is required.
Table 3-3. Minimum Guard Resistances
Range Minimum value for Ra and Rb
2 , 20 200
200 2 k
2 k 20 k, 200 k, 2 M 20 k
20 M, 200 M, 2 G, 20 G 200 k
Providing that Ra and Rb are greater than the values shown in Table 3-3. Minimum
Guard Resistances, and the
Guard resistance (Rg) is less than 1 ; the actual value can
be calculated from the displayed value Rd by:
Rx = Rd x (1 + E)
Deviation fraction ‘E’ can be found within 1 % by the simplified formula:
E = (Rd x Rg) / (Ra x Rb)
(Where Rg is the Guard lead-resistance from the junction of Ra and Rb)
Example:
If Rd = 100 , Rg = 1 , Ra = Rb = 10 k, then the value of E is given by:
E = (100 x 1) / (10 k x 10 k) = 10
-6
(1 ppm of readings)
The value of Rx is thus given by:
Rx = 100 x (1 + 10
-6
) Ohms,
= 100.0001 Ohms
3-19

Table of Contents

Other manuals for Fluke 8508A

Related product manuals