Making Measurements
Using the Measurement Functions
3
HIV
Resolution
Press the Resl key to see the HIRESL menu:
HiVRESL Menu
adj080f.eps
5
6
5-1/2-digits resolution
6-1/2-digits resolution
7
8
7-1/2-digits resolution
8-1/2-digits resolution
Refer to Table 3-2 for the ADC integration times applicable to the various resolution and
Fast On/Off settings.
To revert to the
HiV CONFIG menu, press the Config key.
To revert to the
HiV menu from either HiV RESL or HiV CONFIG, press the HiV
menu key.
HIV
- Movement between Menus
adj085f.eps
HiV Movement Between Menu
True Ohms
adj081f.eps
Tru Menu
Press the PLUS key to see the expanded Ohms selection for True Ohms function and
the High Voltage Ohms function. Press the Tru key for the Tru menu.
The
Trumode takes two measurements per reading. The second measurement is
made with the current reversed relative to the first measurement. The two
measurements are combined to eliminate the effects of any external EMFs that
may be present.
 This function provides only 4-wire measurements of resistance, in decade
ranges fro
m
2 to 20 k. The stimulus current is fed through the test
3-15